missing translation for 'onlineSavingsMsg'
Learn More
Learn More
Descrizione
- Maximum error for probe assembly is ±1.1°C + 0.12% of reading below 300°F, ±0.35% of reading above 300°F.
- Alpha coefficient is 0.003850 omega/omega/°C (DIN IEC 751).
- Resistance at ice point (0°C) is 100 omega.
- Time constant is the time required to reach 63.2% of the new reading, to determine how long it will take to reach 99% of the new reading multiply the time constant by five.
- The temperature range is for the probe tip only.
- The handle material may be affected if the probe is rated higher than the handle depending on exposure time, application, and handle material.
- ABS handles are rated to 220°F (104°C).
- PVC handles are rated to 200°F (93°C).
- Stainless steel handles are rated at 450°F (232°C).
- Mini connectors are rated to 392°F (200°C).
Specifica
Specifica
| Lunghezza sonda (sistema imperiale) | 10 in. |
| Materiale | ABS Plastic, PFA-coated Stainless Steel (316 ) |
| Tipo di sonda | RTD |
| Descrizione | PFA-Ctd RTD Probe, 100 Ohm, ANSI 3-Blade Connector, 10 in. L, 5ft Coil Cord |
| Intervallo di temperatura (sistema imperiale) | -58°F to +400°F |
| Temperature Range (Metric) | -50°C to +204°C |
| Accuracy | ±1.1°C, ±0.12% of reading |
Titolo del prodotto
Facendo clic su Invia, l'utente riconosce che potrebbe essere contattato da Fisher Scientific in merito al feedback fornito in questo modulo. Non condivideremo le vostre informazioni per altri scopi. Tutte le informazioni di contatto fornite saranno conservate in conformità con la nostra Politica sulla privacy. Informativa sulla privacy.
Individuate un'opportunità di miglioramento?