missing translation for 'onlineSavingsMsg'
Learn More
Learn More
Mettler Toledo™ FiveEasy™ Plus FP30 Conductivity Meters
Sfoglia Tutto Mettler Toledo Prodotti
Click to view available options
Tipo:
Conductivy Meter
Meter with Standard Kit
Dimensione della confezione:
1 unità
Descrizione
Includes
LE703 sensor, installation QuickGuide, 1413μS/cm and 12.88mS/cm standard sachets
- Compact, space saving design with ABS housing
- Remove sensor holder after use and store on the side of the meter
- Large, well-structured 4.3 in. segmented LCD display for easy viewing
- Intuitive button arrangement and simple menu for easy operation
- RS232 or USB ports to export data directly either to a printer or a PC for further processing
- One point calibration with three predefined standards, manual cell constant input
- Mini-DIN sensor input
- Features analog output, Automatic and manual endpoint recording, Acoustic endpoint signal, Visual endpoint signal
- Ideal for wide range of laboratory applications
- Conductivity Range: 0.01μS/cm to 500mS/cm
- Conductivity Resolution: Automatic range
- Conductivity Accuracy: ±0.5% of the measured value
- TDS Range: 0.01mg/L to 300g/L
- TDS Resolution: Automatic range
- TDS Accuracy: ±0.5% of the measured value
- Salinity Range: 0.00 to 42 psu
- Salinity Resolution: Automatic range
- Salinity Accuracy: ±0.5% of the measured value
- Temperature Range: -5° to +105°C (23° to 221°F)
- Temperature Resolution: 0.1°C
- Temperature Accuracy: 0.3°C
- Temperature Compensation: Linear: 0.00%/°C to 10.00%/°C, reference temperature: 20° and 25°C
- Memory Size: 200 measurements, current calibration
- Power Supply: 100 to 240V, 50/60Hz, 12 VDC
- Dimensions: 22.7 x 14.7 x 7cm (9 x 5.7 x 2.7 in.)
- Weight: 0.63kg (1.4 lb.)
- FP30 Meter only: Suited for applications that require a specific sensor not included in a kit
- FP30 Standard Kit: Suited for aqueous samples with mid to high conductivity
Specifica
Specifica
| Descrizione | FiveEasy Plus FP30-Std |
| Include | FP30 Meter, LE703 pH sensor, electrode holder, 2 buffer sachets each for pH 1413 mS/cm and 12.88 mS/cm), test certificate |
| Precisione (conduttività) | ±0.5% of the measured value |
| Da utilizzare con (applicazione) | Suited for aqueous samples with mid to high conductivity |
| Measurement Modes | Conductivity |
| Calibrazione | One point calibration with three predefined standards, manual cell constant input |
| Intervallo (conduttività) | 0.01 μS/cm to 500 mS/cm |
| Risoluzione (conduttività) | Automatic range |
| Compensazione della temperatura | Linear: 0.00%/°C to 10.00%/°C, reference temperature: 20° and 25°C |
| Intervallo (salinità) | 0 to 42 psu |