This standard is qualified for use on instruments having a spectral bandwidth of <3nm.
Codice prodotto. 11823654 Fornitore Thermo Scientific™ N. del fornitore 222226400

Please to purchase this item. Need a web account? Register with us today!

Les retours ne sont pas autorisés pour ce produit. Consulta la politica di reso
This standard is qualified for use on instruments having a spectral bandwidth of <3nm.

Thermo Scientific™ NIST™-traceable Holmium Oxide Solution Standard verifies wavelength accuracy of UV-Visible spectrophotometers.

Thermo Scientific NIST™-traceable Holmium Oxide Solution Standard verifies wavelength accuracy of UV-Visible spectrophotometers.

Holmium Oxide in Perchloric Acid Solution is a preferred standard for verifying the wavelength accuracy of a spectrophotometer according to international pharmacopoeia guidelines.
  • Permanently sealed in a quartz cuvette allowing it to be used into the deep UV range
  • Exhibits sharp, stable peaks within the range of 219 to 650nm—can be easily correlated to wavelengths of the peak maxima
  • Readings observed for each peak are verified against expected values on the certificate supplied with the standard
  • Readings not within the expected range may be indicative of an instrument performance error

This standard is qualified for use on instruments having a spectral bandwidth of <3nm.
TRUSTED_SUSTAINABILITY

Specifica

Calibrazione NIST/NPL Traceable
Descrizione Wavelength Standard; Verification Standard
Tipo Holmium Oxide in Perchloric Acid Solution
Titolo del prodotto
Wählen Sie ein Thema

Indem Sie auf Absenden klicken, erklären Sie sich damit einverstanden, dass Fisher Scientific sich mit Ihnen in Verbindung setzen kann, um Ihr Feedback in diesem Formular zu bearbeiten. Wir werden Ihre Informationen nicht für andere Zwecke weitergeben. Alle bereitgestellten Kontaktinformationen werden in Übereinstimmung mit unserer Datenschutzrichtlinie aufbewahrt. Datenschutzrichtlinie.