missing translation for 'onlineSavingsMsg'
Learn More

Thermo Scientific™ (IOPQ) Qualification and Performance Service

Codice prodotto. 30049051 Sfoglia Tutto Thermo Scientific Prodotti
Cambia vista
Click to view available options
Product Type:
IQ/OQ/PQ Service
Dimensione della confezione:
1 unità
1 product options available for selection
Product selection table with 1 available options. Use arrow keys to navigate and Enter or Space to select.
Codice prodotto. Product Type unitSize
30049051 IQ/OQ/PQ Service 1 unità
Use arrow keys to navigate between rows. Press Enter or Space to select a product option. 1 options available.
1 options
Les retours ne sont pas autorisés pour ce produit. Consulta la politica di reso
Codice prodotto. 30049051 Fornitore Thermo Scientific™ N. del fornitore IOPQPCKE89003711

Please to purchase this item. Need a web account? Register with us today!

Les retours ne sont pas autorisés pour ce produit. Consulta la politica di reso

Qualification service for New Heratherm Chamber 750 L (including documentation and 1 temp mapping).

  • IQ: Verifies that the installation site and equipment environment meet manufacturer-specified environmental requirements.
  • Verifies that the equipment, manuals, supplies and any other accessories arrived undamaged as specified in the sales order.
  • Verifies equipment and any other accessories are assembled and installed.
  • Configuration information for each system component gets recorded. Installation of Equipment is NOT Included.
  • OQ: Verifies important equipment functions and ensures that the equipment operates as expected by the manufacturer and in conformance to standards and requirements.
  • Key test: Door Ajar, Power Failure, Temperature Alerts PT: Temperature Mapping protocol performed at customer specified set point with provisions up to 15 measurement points, plus 1 ambient measurement point.
  • Temperature measurements will be performed in air or with simulated product loaded in chamber with a data collection interval of 1 minute for up to 24 hours, or configurable to customer needs.
  • Study requirements greater than 24 hours will be reviewed and quoted.
  • Temp Mapping service will include: (a) 1 open door temperature recovery test (b)1 power failure test (c) Min and max values along with averages (d) Temperature graph over a time period (e) Detailed reports will include raw probe data. All of this will be performed and delivered by a Thermo Fisher Qualified Engineer equipped with training certificates,temperature mapping data logger calibrated annually with metrology certificates.
TRUSTED_SUSTAINABILITY

Specifica

Product Type IQ/OQ/PQ Service
Titolo del prodotto
Selezionare un problema

Facendo clic su Invia, l'utente riconosce che potrebbe essere contattato da Fisher Scientific in merito al feedback fornito in questo modulo. Non condivideremo le vostre informazioni per altri scopi. Tutte le informazioni di contatto fornite saranno conservate in conformità con la nostra Politica sulla privacy. Informativa sulla privacy.