Thermo Scientific™ NIST™ traceable standards verify spectrophotometer stray light specification.
Each Thermo Scientific Stray Light Standard stops transmitting light below a specific wavelength. Below this cut-off, any measured value can be attributed to instrumental stray light. A measured value of stray light that does not meet instrument specifications may indicate a problem with the instrument lamp and may result in analytical errors.
The following standards are available for stray light measurements:
- Potassium Chloride, cutoff at 200nm
- Sodium Iodide, cutoff at 260nm
- Sodium Nitrate, cutoff at 390nm
Stray Light Standards are provided in permanently sealed quartz cuvettes. A water blank corresponding to each specific standard is supplied with every Stray Light Standard for performing blank readings.
Spezifikation
Spezifikation
| Calibrazione | NIST Traceable |
| Tipo | Stray Light Standard, NaI |
| Wavelength Range | 250 to 270 nm |
| Da utilizzare con (apparecchiatura) | Evolution 60S, 201, 220 260 Bio, 300, GENESYS 10S, BioMate 3S |
| Include | Blank and 10 g/L NaI Standard |
Indem Sie auf Absenden klicken, erklären Sie sich damit einverstanden, dass Fisher Scientific sich mit Ihnen in Verbindung setzen kann, um Ihr Feedback in diesem Formular zu bearbeiten. Wir werden Ihre Informationen nicht für andere Zwecke weitergeben. Alle bereitgestellten Kontaktinformationen werden in Übereinstimmung mit unserer Datenschutzrichtlinie aufbewahrt. Datenschutzrichtlinie.